HIAC PODS+ Online Mode & Filter Cart Mode
Joe Dabbs & Bill F. Bars | Beckman Coulter, Inc., 481 California Ave Grants Pass, OR 97526 USAAbstract
The HIAC PODS+ has three different sampling modes, Bottle Mode, Online Mode and Filter Cart Mode. The Bottle Mode is for sampling fluid that is contained within a bottle and then the bottle is placed in the sample cup chamber. Online Mode is when you connect the HIAC PODS+ directly to a system and the instrument makes measurements periodically. Filter Cart Mode is identical to Online Mode but with added features, like controlling an external device or simply turning a light on when the fluid becomes dirty. Knowing the differences between Online Mode and Filter Cart Mode will allow operators to get the most out of their PODS+.
Difference between Online Mode and Filter Cart Mode
Both are online configurations of the PODS+ instrument where the instrument is connected directly to an active fluid system:
Online Mode
Makes measurements periodically from an active system to determine cleanliness in a dynamic environment.

Filter Cart Mode
Identical to Online Mode with the exception of the provision to specify a desired channel count level or cleanliness code at a point the operator would like to control an external device when that count or code is reached. This can be utilized for both “Clean to Dirty” or “Dirty to Clean”, scenarios.

*Waste line must have little or no pressure for fluid to properly flow through the PODS+
Scenario 1 – Filter Cart Mode Clean to Dirty
Connect the PODS+ to an active system and taking a particle count measurement hourly. Like most systems, contamination levels increase steadily while the system is in operation. The operator would like to know either visually or audibly without constantly monitoring the instrument when the contamination level reaches an ISO code of 20 on the 4µm channel. The PODS+ can be connected in the online manner to the system to perform the periodic sample testing. The Filter Cart Mode is selected and programmed to trigger an external device when the system reaches the desired code of 20 on the 4µm channel (Clean to Dirty option). Once the desired code is reached, the PODS+ is equipped with an I/O control port with relay contact control to drive an alarm light, audible alarm, or other electronic device to signal the operator the system is dirty and in need of service.
Scenario 2 – Filter Cart Mode Dirty to Clean
This is a follow onto the above scenario where now you have a dirty system that needs to be connected to a filter cart for cleaning. The PODS+ can be configured with the “Dirty to Clean” option and a cleanliness trigger level of an ISO code of 14 on the 4µm channel. Start the system and start the PODS+ sampling. The I/O port would be connected to control the filter cart pump so that when the cleanliness code of 14 on the 4µm channel is reached, and the system is sufficiently clean, the control signal can turn off the filter cart pump. This is a walk away scenario where the instrument, filter cart, and system are started simultaneously and when the process is complete the Filter cart pump is shut down to be disconnected and removed at the customer’s convenience.
Hardware
For both the Online and Filter Cart mode the following accessories are required. Online Adapter Kit Part number: B95366 Included in the Kit: On-Line Adapter, High Pressure Hose Mini-mess/SAE J514 40”, High Pressure Hose Mini-mess/Mini-mess 78”, Drain Tubing 6’, antistatic PTFE with metal quick connect, and a FFKM O-ring.
Helpful Links
-
Notas de aplicación
- 17-Marker, 18-Color Human Blood Phenotyping Made Easy with Flow Cytometry
- 21 CFR Part 11 Data Integrity for On-line WFI Instruments
- 8011+ Reporting Standards Feature and Synopsis
- Achieving Compliant Batch Release – Sterile Parenteral Quality Control
- Air Particle Monitoring ISO 21501-4 Impact
- An Analytical Revolution: Introducing the Next Generation Optima AUC
- Analyzing Mussel/Mollusk Propagation using the Multisizer 4e Coulter Counter
- Automated 3D Cell Culture and Screening by Imaging and Flow Cytometry
- Automated Cell Plating and Growth Assays
- Automated Cell Transfection and Reporter Gene Assay
- Mediciones automatizadas de concentración y viabilidad de las células sanguíneas del cordón umbilical utilizando el VI-CELL XR
- Automated Genomic Sample Prep-Ampure XP PCR
- Automated Genomic Sample Prep-RNAdvance
- Automated Genomic Sample Prep-Sample Quality Control
- Automated salt-assisted liquid-liquid extraction
- Automated Sample Preparation for the Monitoring of Pharmaceutical and Illicit Drugs by LC-MS/MS
- Métodos de transfección automatizados
- Automated XTT Assay for Cell Viability Analysis
- Automatización de un gradiente de densidad lineal para purificación de un complejo proteína-ligandos
- Automatización del control de calidad biofarmacéutica para reducir costes y mejorar la integridad de los datos
- Automatización de los ensayos Bradford
- Automatización del desarrollo de líneas celulares
- Automating Cell Line Development
- Automatización de procesos basados en células
- Automatización del flujo de trabajo de desarrollo de líneas celulares
- La nueva centrífuga Avanti J-15 mejora la protección de las muestras y maximiza la recuperación de muestras
- El nuevo perfil de desaceleración de ahorro de tiempo de la centrífuga Avanti J-15 mejora la eficiencia del flujo de trabajo
- Avanti JXN Protein Purification Workflow
- Evite los errores al automatizar el recuento de viabilidad celular para el control de calidad biofarmacéutico
- Beer, Evaluation of Final Product and Filtration Efficiency
- La preparación automatizada de muestras genómicas de Biomek acelera la investigación
- Biomek Automated NGS Solutions acelera la investigación genómica
- Preparation and purification of carbon nanotubes using an ultracentrifuge and automatic dispensing apparatus, and analysis using an analytical centrifuge system
- Recuento de células del analizador de viabilidad celular Vi–Cell BLU
- Cell Line Development – Data Handling
- Cell Line Development – Hit Picking
- Cell Line Development – Limiting Dilution
- Cell Line Development – Selection and Enrichment
- Cambios en las reclasificaciones de BPF de sala limpia en vigor
- Caracterizando la insulina como producto biofarmacéutico usando el ultracentrifugado analítico
- Clean Cabinet Air Particle Evaluation
- Monitorización ambiental rutinaria de sala limpia – Guía de la FDA sobre 21 CFR Parte 11 Integridad de los datos
- Comparación de la calidad de los datos y la resolución óptica del Optima AUC de nueva generación con el ProteomeLab de eficacia demostrada en un sistema de proteínas modelo
- Realización de la prueba de recuperación de sala limpia ISO 14644-3 con el MET ONE 3400
- Considerations of Cell Counting Analysis when using Different Types of Cells
- Consistent Cell Maintenance and Plating through Automation
- Normas de control y recomendaciones de métodos para el LS 13 320 XR
- Eficacia del recuento: Contadores de partículas de aire MET ONE y cumplimiento de ISO-21501
- Critical Particle Size Distribution for Cement using Laser Diffraction
- CytoFLEX
- Detección de humedad en fluidos hidráulicos, aceite y combustibles
- Determination of Size and Concentration of Particles in Oils
- El aislamiento eficiente de ácidos nucleicos sin kit utiliza una combinación de métodos de separación mediante precipitación y centrifugación
- dsDNA Quantification with the Echo 525 Liquid Handler for Miniaturized Reaction Volumes, Reduced Sample Input, and Cost Savings
- Configuración de compensación para reactivos DURAClone de alto contenido
- Echo System-Enhanced SMART-Seq v2 for RNA Sequencing
- Efficient Factorial Optimization of Transfection Conditions
- Enhancing Vaccine Development and Production
- Enumeration And Size Distribution Of Yeast Cells In The Brewing Industry
- Farmacopea europea EP 2.2.44 y carbono orgánico total
- Evaluación del rendimiento de instrumento a instrumento del analizador de viabilidad celular Vi-CELL BLU
- FBS mermado de exosomas usando la centrifugación Beckman Coulter: La opción rentable y uniforme
- Flexible ELISA automation with the Biomek i5 Workstation
- Automatización completa del flujo de trabajo SISCAPA® utilizando una estación de trabajo de automatización de laboratorio Biomek NXP
- Análisis celular totalmente automatizado mediante citometría de flujo
- Get Control in GMP Environments
- g-Max: Added Capabilities to Beckman Coulter's versatile Ultracentrifuge Line
- Grading of nanocellulose using a centrifuge
- A method of grading nanoparticles using ultracentrifugation in order to determine the accurate particle diameter
- Grading of pigment ink and measurement of particle diameter using ultracentrifugation / dynamic light scattering
- HIAC Industrial: nuestra solución general para la prueba de fluidos de potencia para todas las aplicaciones
- HIAC PODS+ Online Mode & Filter Cart Mode
- HIAC PODS+ versus Parker ACM-20 Performance comparison
- A complete workflow for high-throughput isolation of DNA and RNA from FFPE samples using Formapure XL Total on the KingFisher™ Sample Purification System: an application for robust and scalable cancer research and biomarker discovery
- A Highly Consistent BCA Assay on Biomek i-Series
- A Highly Consistent Bradford Assay on Biomek i-Series
- A Highly Consistent Lowry Method on Biomek i-Series
- Highly Reproducible Automated Proteomics Sample Preparation on Biomek i-Series
- Desarrollo de líneas celulares: selección de coincidencias
- Cómo utilizar la dispersión lateral violeta (V-SSC) para detectar nanopartículas en el citómetro de flujo CytoFLEX
- Cómo la dispersión lateral violeta permite la detección de nanopartículas
- HRLD Recommended Volume Setting
- Hydraulic Oil Measurement
- ICH Q2: el desafío de medir el carbono orgánico total en sistemas de agua farmacéutica modernos
- ICH Q2 – The Challenge of Measuring Total Organic Carbon in Modern Pharmaceutical Water Systems
- Importancia de la medición del COT en el API (agua para inyección) a la luz del cambio de la Farmacopea Europea
- Dependencia con la temperatura del radio hidrodinámico de una proteína intrínsecamente desordenada medida en la ultracentrífuga analítica Optima AUC.
- Isolation of cell-free DNA (cfDNA) from plasma using Apostle MiniMax™ High Efficiency cfDNA Isolation kit— comparison of fully automated, semi-automated and manual workflow processing
- Problemas con las pruebas de contaminación de los combustibles para aviones
- Aprovechamiento del Vi-CELL MetAfLEX para monitorizar la actividad metabólica celular
- Linealidad de BSA utilizando ópticas de absorbancia y interferencia
- Long Life Lasers
- LS 13 320 XR: Sample Preparation - How to measure success
- Particle Size Analysis Simple, Effective and Precise
- Matching Cell Counts between Vi–CELL XR and Vi–CELL BLU
- MET ONE Sensor Verification
- Metal colloid purification and concentration using ultracentrifugation
- Separation and purification of metal nanorods using density gradient centrifugation
- Método para determinar el ajuste del parámetro de tipo de célula para que concuerde con el Vi CELL XR heredado
- Minimal Sample to Sample Carry Over with the HIAC 8011+
- Minimizar la variabilidad del proceso en la producción de agua potable embotellada
- Tendencias modernas en la monitorización de partículas no viables durante el procesamiento aséptico
- Ultracentrifugado analítico de múltiples longitudes de onda de albúmina de suero humano complejada con porfirina
- Particle diameter measurement of a nanoparticle composite - Using density gradient ultracentrifugation and dynamic light scattering
- Qué se hace ahora con el ACFTD discontinuado
- Optimizing the HIAC 8011+ Particle Counter for Analyzing Viscous Fluids
- Optimizing the Multisizer 4e Coutler Counter for use with Small Apertures
- Optimización del flujo de trabajo Eficiencia de la monitorización ambiental rutinaria de sala limpia
- Recuento de partículas en aplicaciones mineras
- Pruebas de partículas en líneas de gas a alta presión de sala limpia en conformidad con ISO 14644 fáciles de realizar con las calibraciones de gas del MET ONE 3400
- Pharma Manufacturing Environmental Monitoring
- Pharma Manufacturing Paperless Monitoring
- Principles of Continuous Flow Centrifugation
- Protein purification workflow
- Background Subtraction
- Calibrating the QbD1200 TOC Analyzer
- Detection Limit
- Inorganic Carbon Removal
- JP SDBS Validation
- Method Overview
- Overload Recovery
- QbD1200 Preparing Reagent Solution
- USP System Suitability
- Registros electrónicos de control de calidad para el cumplimiento de 21 CFR parte 11
- Using the Coulter Principle to Quantify Particles in an Electrolytic Solution for Copper Acid Plating
- RCC A standardized automated approach for exosome isolation and characterization
- RCC Avanti J 15 Series Maximizing Sample Protection and Sample Recovery
- RCC Avanti J 15 Series Time Savings and Workflow Efficiency
- Investigaciones de causa raíz para sistemas de agua farmacéutica
- Purificación de plásmidos escalable con CosMCPrep
- Comparación de especificaciones del Vi–CELL XR y el Vi–CELL BLU
- Especificación de la monitorización de partículas no viables para procesamiento aséptico
- Un enfoque estandarizado y automatizado para el aislamiento y la caracterización de exosomas utilizando instrumentos Beckman Coulter
- Cambio de prueba de aceite a agua y viceversa utilizando el HIAC 8011+ y el HIAC PODS+
- Análisis avanzado de subconjuntos de linfocitos T humanos en el citómetro de flujo CytoFLEX utilizando un reactivo seco DURAClone de 13 colores en tubo
- Using k-Factor to Compare Rotor Efficiency
- USP 787 Small Volume Testing
- Validación de los analizadores de carbono orgánico total en línea para pruebas de liberación utilizando ICH Q2
- Descontaminación con peróxido de hidrógeno vaporizado del instrumento Vi–CELL BLU
- Vi CELL BLU FAST Mode Option
- Cumplimiento reglamentario del Vi-CELL BLUE: 21 CFR parte 11
- A workflow for medium-throughput isolation of cfDNA from plasma samples using Apostle MiniMax™ on the KingFisher™ Technology