Considerations of Cell Counting Analysis when using Different Types of Cells
Beckman Coulter Life Sciences is proud to introduce our new Vi-CELL BLU Cell Viability Analyzer. The Vi-CELL BLU leverages the key performance features of the Vi-CELL XR but incorporates many design improvements that our customers have requested over the years.
While a seemingly straightforward application, automated cell counting can be influenced by a variety of conditions and variables arising from both the sample and instrument. With the new Vi-CELL BLU we have introduced the capability to load 96 samples into a standard plate and run this as a single experiment. However this requires approximately 3 hours to run the entire plate with may not be ideal, particularly for cell types that generally have low viability or are prone deterioration once outside the incubator or bioreactor.
The plate loader also allows us to assess multiple different experimental conditions in a single run thus providing the opportunity to evaluate a range of conditions or criteria.
To demonstrate these capabilities we reviewed several different cell types and cell preparation approaches using standard cells.
Cell Counting Analysis
Cells cultures of CHO (Chinese Hamster Ovary), EL4 (Mouse T Lymphocyte Cell Line), SF9 (Ovarian tissue from Spodoptera frugiperda) insect cells and of HELA cells were used to assess the effects of long analysis durations and different sample preparation conditions on different types of cells.
CHO Cell Sample Preparation Impact
Centrifugation and resuspension are routine methods for washing and concentrating cells ahead of reseeding or use in other experiments. CHO cells are generally considered to quite durable, but they do have very particular growth requirements and require specialized media for best health. In the experiment below the cells were centrifuged at 1000 g for 5 minutes, a relatively mild spin before being resuspended in media or buffer to assess the impact, if any, of cell preparation on cell health
Results
Treatment | Concentration (M/mL) | Viability (%) | Diameter (µm) |
Untreated | 4.83 | 94.30 | 16.12 |
1 spin resuspend in media | 5.05 | 94.48 | 16.17 |
1 spin resuspend in PBS | 4.70 | 65.69 | 14.17 |
2 spins resuspend in PBS | 4.45 | 53.77 | 13.99 |
The results show that centrifugation and resuspension in recommended media has no appreciable impact on the CHO cells. However resuspension in PBS buffer results in a drastic loss of viability and significant reduction in average cell diameter. There is also some minor reduction in cell concentration which may be expected from losses during resuspension. Spinning and resuspending a second time in PBS has a further negative impact on cell viability and cell size


Figure 2. Impact of sample preparation method on CHO cells
EL4 Cells Mixing Cycle Analysis
EL4 cells are generally considered to be fragile and even under ideal conditions grow slowly and show viability <70%. EL4 cells were prepared and dispensed into a 96 well plate. Cell Type profiles were created using the standard Mammalian cell profile as source and differing levels of mixing cycles. The goal was to assess the impact, if any, of increased agitation on a cell type known to be easily damaged by vortex mixing and centrifugation
Results
The results for the EL4 cells are given below:
Mix Cycles | Cell Count | Viable Cell Count | Concentration (M/mL) | Viability (%) | Diameter (µm) |
1 | 4648 | 2837 | 1.72 | 61.06 | 11.34 |
3 | 4637 | 2715 | 1.72 | 58.54 | 11.34 |
5 | 4660 | 2582 | 1.73 | 55.69 | 11.34 |
7 | 4660 | 2448 | 1.73 | 52.53 | 11.32 |
While the cell counts, concentration and diameter values show no statistical significance there is a marked drop in viable cell count and consequently in cell viability with increased mixing cycles across all dilutions. This may indicate that a Cell Type Profile with reduced aspiration and mixing cycles may be more suitable for this type of cell and excessive agitation may not be ideal and should be avoided
Figure 3. Impact of mixing cycles on EL4 cells
SF9 Insect Cells
Even when grown under ideal conditions SF9 cells do not achieve high densities easily and typically have viability below 60%. It is therefore possible that the prolonged instrument time required for a plate run may negatively impact the cell culture. To assess this SF9 cells were plated and run using the standard Insect Cell profile.
The results show that over the course of the 3 hour run there was no detectable impact on cell viability and no notable change in cell diameter that may occur if the cells are dying or swelling. As such it would appear this cell line, although low in viability, is robust enough to spend extended periods outside the cell incubator environment without adverse effects
Figure 4. Viability and average diameter of SF9 cells over time. No significant change noted over course of experiment.
HELA Cell Cultures
HELA cells were grown to a high state of confluence in order to stress the cells. The cells were then harvested and dispersed into suspension. The cells were then plated out on a 96 well plate and data collected using standard Mammalian cell profile. The plate was run collecting data by columns to ensure the same time interval between samples in the rows.
It was noted that the later samples in the plate run were reporting slightly higher concentrations. Further investigations showed that the cells were increasing in average diameter as time progressed.
Figure 5. Percentage increase in average diameter of HELA cells over time
Viewing the acquired images shows an increasing number of cells which appear to be swelling or blebbing along with a large number of non-cellular vesicles. While the latter do not usually impact the cell count, depending on the cell profile settings, they could be counted as small cells. The swelling effect causes the average cell diameter to increase almost 10% over the course of the 3 hours it takes to run the plate and may cause some smaller cells just below the lower diameter threshold to become included in the cell counts.
Figure 6. Cell swelling and blebbing in HELA cells after >2hrs outside incubator
Conclusions
It is recommended that prior to running a cell culture on a complete plate that some initial evaluation is performed to determine the robustness of the cells involved and their ability to tolerate the sample preparation methods and prolonged exposure outside the incubator environment. Even engineered cells such as CHO cells or immortal HELA cells, which are generally selected to be quite durable, can be effect by stress or particular sample preparation conditions.
Helpful Links
-
Material de lectura
-
Notas de aplicación
- 17-Marker, 18-Color Human Blood Phenotyping Made Easy with Flow Cytometry
- 21 CFR Part 11 Data Integrity for On-line WFI Instruments
- 8011+ Reporting Standards Feature and Synopsis
- Achieving Compliant Batch Release – Sterile Parenteral Quality Control
- Air Particle Monitoring ISO 21501-4 Impact
- An Analytical Revolution: Introducing the Next Generation Optima AUC
- Analyzing Mussel/Mollusk Propagation using the Multisizer 4e Coulter Counter
- Automated 3D Cell Culture and Screening by Imaging and Flow Cytometry
- Automated Cell Plating and Growth Assays
- Automated Cell Transfection and Reporter Gene Assay
- Mediciones automatizadas de concentración y viabilidad de las células sanguíneas del cordón umbilical utilizando el VI-CELL XR
- Automated Genomic Sample Prep RNAdvance
- Automated Genomic Sample Prep Sample Quality Control
- Automated salt-assisted liquid-liquid extraction
- Automated Sample Preparation for the Monitoring of Pharmaceutical and Illicit Drugs by LC-MS/MS
- Métodos de transfección automatizados
- Automated XTT Assay for Cell Viability Analysis
- Automatización de un gradiente de densidad lineal para purificación de un complejo proteína-ligandos
- Automatización del control de calidad biofarmacéutica para reducir costes y mejorar la integridad de los datos
- Automatización de los ensayos Bradford
- Automatización de procesos basados en células
- Automating Cell Line Development
- La nueva centrífuga Avanti J-15 mejora la protección de las muestras y maximiza la recuperación de muestras
- El nuevo perfil de desaceleración de ahorro de tiempo de la centrífuga Avanti J-15 mejora la eficiencia del flujo de trabajo
- Avanti JXN Protein Purification Workflow
- Evite los errores al automatizar el recuento de viabilidad celular para el control de calidad biofarmacéutico
- Beer, Evaluation of Final Product and Filtration Efficiency
- La preparación automatizada de muestras genómicas de Biomek acelera la investigación
- Biomek Automated NGS Solutions Accelerate Genomic Research
- Biomek i-Series Automated AmpliSeq for Illumina® Library Prep Kit
- Biomek i-Series Automated Beckman Coulter Agencourt RNAdvance Blood Kit
- Biomek i-Series Automated Beckman Coulter Agencourt RNAdvance Cell
- Biomek i-Series Automated Beckman Coulter Agencourt SPRIselect for DNA Size Selection
- Biomek i-Series Automated Beckman Coulter AMPure XP PCR Purification System
- Biomek i-Series Automated IDT® xGen Hybridization Capture of DNA libraries on Biomek i7 Hybrid Genomics Workstation
- Biomek i-Series Automated Illumina Nextera DNA Flex Library Prep Kit
- Biomek i-Series Automated Illumina® Nextera XT DNA Library Prep Kit
- Biomek i-Series Automated Illumina TruSeq DNA PCR-Free Library Prep Kit
- Biomek i-Series Automated Illumina TruSeq® Nano DNA Library Prep Kit
- Biomek i-Series Automated Illumina TruSeq® Stranded mRNA Sample Preparation Kit Protocol
- Biomek i-Series Automated Illumina TruSeq® Stranded Total RNA Sample Preparation Kit Protocol
- Biomek i–Series Automated Illumina® TruSight Tumor 170 32 Sample Method
- Biomek i-Series Automated KAPA HyperPrep and HyperPlus Workflows
- Biomek i-Series Automated New England Biolabs NEBNext® Ultra IITM DNA Library Prep Kit
- Biomek i-Series Automated SurePlex PCR and VeriSeq PGS Library Prep for Illumina®
- Biomek i-Series Automation of the Beckman Coulter GenFind V3 Blood and Serum DNA Isolation Kit
- Biomek i-Series Automation of the Apostle MiniMax™ High Efficiency cfDNA Isolation Kit
- Biomek i-Series Automation of the Beckman Coulter Agencourt DNAdvance Genomic DNA isolation Kit
- Preparation and purification of carbon nanotubes using an ultracentrifuge and automatic dispensing apparatus, and analysis using an analytical centrifuge system
- Recuento de células del analizador de viabilidad celular Vi–Cell BLU
- Viability Assessment of Cell Cultures Using the CytoFLEX
- Cell Line Development – Data Handling
- Cell Line Development – Hit Picking
- Cell Line Development – Limiting Dilution
- Cell Line Development – Selection and Enrichment
- Cellular Analysis using the Coulter Principle
- cfDNA Extraction Efficiency Affects NGS Data
- Cambios en las reclasificaciones de BPF de sala limpia en vigor
- Caracterizando la insulina como producto biofarmacéutico usando el ultracentrifugado analítico
- Clean Cabinet Air Particle Evaluation
- Monitorización ambiental rutinaria de sala limpia – Guía de la FDA sobre 21 CFR Parte 11 Integridad de los datos
- Comparing Data Quality & Optical Resolution of the Next Generation Optima AUC to the Proven ProteomeLab on a Model Protein System
- Realización de la prueba de recuperación de sala limpia ISO 14644-3 con el MET ONE 3400
- Considerations of Cell Counting Analysis when using Different Types of Cells
- Consistent Cell Maintenance and Plating through Automation
- Normas de control y recomendaciones de métodos para el LS 13 320 XR
- Eficacia del recuento: Contadores de partículas de aire MET ONE y cumplimiento de ISO-21501
- Critical Particle Size Distribution for Cement using Laser Diffraction
- Use Machine Learning Algorithms to Explore the Potential of Your High Dimensional Flow Cytometry Data Example of a 20–color Panel on CytoFLEX LX
- CytoFLEX
- Detecting and counting bacteria with the CytoFLEX research flow cytometer: II-Characterization of a variety of gram-positive bacteria
- Detecting Moisture in Hydraulic Fluid, Oil and Fuels
- Determination of Size and Concentration of Particles in Oils
- El aislamiento eficiente de ácidos nucleicos sin kit utiliza una combinación de métodos de separación mediante precipitación y centrifugación
- dsDNA Quantification with the Echo 525 Liquid Handler for Miniaturized Reaction Volumes, Reduced Sample Input, and Cost Savings
- Configuración de compensación para reactivos DURAClone de alto contenido
- Echo System-Enhanced SMART-Seq v2 for RNA Sequencing
- Effective Miniaturization of Illumina Nextera XT Library Prep for Multiplexed Whole Genome Sequencing and Microbiome Applications
- Efficient Factorial Optimization of Transfection Conditions
- Enhancing Vaccine Development and Production
- Enumeration And Size Distribution Of Yeast Cells In The Brewing Industry
- Farmacopea europea EP 2.2.44 y carbono orgánico total
- Evaluación del rendimiento de instrumento a instrumento del analizador de viabilidad celular Vi-CELL BLU
- FBS mermado de exosomas usando la centrifugación Beckman Coulter: La opción rentable y uniforme
- Flexible ELISA automation with the Biomek i5 Workstation
- Análisis celular totalmente automatizado mediante citometría de flujo
- Get Control in GMP Environments
- g-Max: Added Capabilities to Beckman Coulter's versatile Ultracentrifuge Line
- Grading of nanocellulose using a centrifuge
- A method of grading nanoparticles using ultracentrifugation in order to determine the accurate particle diameter
- Grading of pigment ink and measurement of particle diameter using ultracentrifugation / dynamic light scattering
- HIAC Industrial: nuestra solución general para la prueba de fluidos de potencia para todas las aplicaciones
- HIAC PODS+ Online Mode & Filter Cart Mode
- HIAC PODS+ versus Parker ACM-20 Performance comparison
- A complete workflow for high-throughput isolation of DNA and RNA from FFPE samples using Formapure XL Total on the KingFisher™ Sample Purification System: an application for robust and scalable cancer research and biomarker discovery
- High-Throughput qPCR and RT-qPCR Workflows
- A Highly Consistent BCA Assay on Biomek i-Series
- A Highly Consistent Bradford Assay on Biomek i-Series
- A Highly Consistent Lowry Method on Biomek i-Series
- Highly Reproducible Automated Proteomics Sample Preparation on Biomek i-Series
- Desarrollo de líneas celulares: selección de coincidencias
- Cómo utilizar la dispersión lateral violeta (V-SSC) para detectar nanopartículas en el citómetro de flujo CytoFLEX
- Cómo la dispersión lateral violeta permite la detección de nanopartículas
- HRLD Recommended Volume Setting
- Automatización del flujo de trabajo de desarrollo de líneas celulares
- Hydraulic Oil Measurement
- ICH Q2: el desafío de medir el carbono orgánico total en sistemas de agua farmacéutica modernos
- ICH Q2 – The Challenge of Measuring Total Organic Carbon in Modern Pharmaceutical Water Systems
- Illumina Nextera Flex for Enrichment on the Biomek i7 Hybrid Genomics Workstation
- Illumina TruSight Oncology 500 Automated on the Biomek NXᴾ Span-8 Genomics Workstation
- Importancia de la medición del COT en el API (agua para inyección) a la luz del cambio de la Farmacopea Europea
- Dependencia con la temperatura del radio hidrodinámico de una proteína intrínsecamente desordenada medida en la ultracentrífuga analítica Optima AUC.
- Isolation of cell-free DNA (cfDNA) from plasma using Apostle MiniMax™ High Efficiency cfDNA Isolation kit— comparison of fully automated, semi-automated and manual workflow processing
- Problemas con las pruebas de contaminación de los combustibles para aviones
- Aprovechamiento del Vi-CELL MetAfLEX para monitorizar la actividad metabólica celular
- Linealidad de BSA utilizando ópticas de absorbancia y interferencia
- Long Life Lasers
- LS 13 320 XR: Sample Preparation - How to measure success
- Particle Size Analysis Simple, Effective and Precise
- Beckman’s LS 13 320 XR Vs. Malvern Mastersizer
- Utilización de algoritmos de aprendizaje automático para la obtención de información detallada sobre la composición de subpoblaciones celulares
- Flow Cytometric Analysis of auto-fluorescent cells found in the marine demosponge Clathria prolifera
- Matching Cell Counts between Vi–CELL XR and Vi–CELL BLU
- MET ONE 3400
- MET ONE Sensor Verification
- Metal colloid purification and concentration using ultracentrifugation
- Separation and purification of metal nanorods using density gradient centrifugation
- Method for Determining Cell Type Parameter Adjustment to Match Legacy Vi CELL XR
- High-throughput Miniaturization of Cytochrome P450 Time-dependent Inhibition Screening Using the Echo 525 Liquid Handler
- Miniaturization and Rapid Processing of TXTL Reactions Using Acoustic Liquid Handling
- Miniaturized Enzymatic Assays with Glycerol
- Miniaturized and High-Throughput Metabolic Stability Assay Enabled by the Echo Liquid Handler
- Miniaturized Multi-Piece DNA Assembly Using the Echo 525 Liquid Handler
- Miniaturized Sequencing Workflows for Microbiome and Metagenomic Studies
- Minimal Sample to Sample Carry Over with the HIAC 8011+
- Minimizar la variabilidad del proceso en la producción de agua potable embotellada
- Tendencias modernas en la monitorización de partículas no viables durante el procesamiento aséptico
- Modular DNA Assembly of PIK3CA Using Acoustic Liquid Transfer in Nanoliter Volumes
- Multi-Wavelength Analytical Ultracentrifugation of Human Serum Albumin complexed with Porphyrin
- Nanoliter Scale DNA Assembly Utilizing the NEBuilder HiFi Cloning Kit with the Echo 525 Liquid Handler
- Particle diameter measurement of a nanoparticle composite - Using density gradient ultracentrifugation and dynamic light scattering
- Identification of Circulating Myeloid Cell Populations in NLRP3 Null Mice
- Qué se hace ahora con el ACFTD discontinuado
- Optimizing the HIAC 8011+ Particle Counter for Analyzing Viscous Fluids
- Optimizing the Multisizer 4e Coutler Counter for use with Small Apertures
- Optimización del flujo de trabajo Eficiencia de la monitorización ambiental rutinaria de sala limpia
- Recuento de partículas en aplicaciones mineras
- Pruebas de partículas en líneas de gas a alta presión de sala limpia en conformidad con ISO 14644 fáciles de realizar con las calibraciones de gas del MET ONE 3400
- PCR Reaction Setup and AMPure XP Application
- PCR Reaction Setup Application
- Pharma Manufacturing Environmental Monitoring
- Pharma Manufacturing Paperless Monitoring
- Analysis of plant genome sizes using flow cytometry: a case study demonstrating dynamic range and measurement linearity
- Principles of Continuous Flow Centrifugation
- Flow Cytometric Approach to Probiotic Cell Counting and Analysis
- Protein purification workflow
- Background Subtraction
- Calibrating the QbD1200 TOC Analyzer
- Detection Limit
- Inorganic Carbon Removal
- JP SDBS Validation
- Method Overview
- Overload Recovery
- QbD1200 Preparing Reagent Solution
- USP System Suitability
- Registros electrónicos de control de calidad para el cumplimiento de 21 CFR parte 11
- Using the Coulter Principle to Quantify Particles in an Electrolytic Solution for Copper Acid Plating
- A Rapid Flow Cytometry Data Analysis Workflow Using Machine Learning- Assisted Analysis to Facilitate Identifying Treatment- Induced Changes
- Investigaciones de causa raíz para sistemas de agua farmacéutica
- Purificación de plásmidos escalable con CosMCPrep
- Automatización completa del flujo de trabajo SISCAPA® utilizando una estación de trabajo de automatización de laboratorio Biomek NXP
- Comparación de especificaciones del Vi–CELL XR y el Vi–CELL BLU
- Especificación de la monitorización de partículas no viables para procesamiento aséptico
- Un enfoque estandarizado y automatizado para el aislamiento y la caracterización de exosomas utilizando instrumentos Beckman Coulter
- Streamlined Synthetic Biology with Acoustic Liquid Handling
- Cambio de prueba de aceite a agua y viceversa utilizando el HIAC 8011+ y el HIAC PODS+
- Análisis avanzado de subconjuntos de linfocitos T humanos en el citómetro de flujo CytoFLEX utilizando un reactivo seco DURAClone de 13 colores en tubo
- Using k-Factor to Compare Rotor Efficiency
- Validación de los analizadores de carbono orgánico total en línea para pruebas de liberación utilizando ICH Q2
- Descontaminación con peróxido de hidrógeno vaporizado del instrumento Vi–CELL BLU
- Vesicle Flow Cytometry with the CytoFLEX
- Vi CELL BLU FAST Mode Option
- Cumplimiento reglamentario del Vi-CELL BLUE: 21 CFR parte 11
- A workflow for medium-throughput isolation of cfDNA from plasma samples using Apostle MiniMax™ on the KingFisher™ Technology
- Automated Research Flow Cytometry Workflow Using DURA Innovations Dry Reagent Technology with the *Biomek i7 Automated Workstation and *CytoFLEX LX Flow Cytometer
- Biomek i7 Hybrid Automated KAPA mRNA HyperPrep Workflow
- Fully Automated Peptide Desalting for Liquid Chromatography–Tandem Mass Spectrometry Analysis Using Beckman Coulter Biomek i7 Hybrid Workstation
- High-throughput Miniaturization of Cytochrome P450 Time-dependent Inhibition Screening Using the Echo 525 Liquid Handler
- Miniaturization of an Epigenetic AlphaLISA Assay with the Echo Liquid Handler and the BMG LABTECH PHERAstar FS
- Miniaturization of Cytochrome P450 Time-dependent Inhibition Screening Using the Echo 555 Liquid Handler
- Utilization of the MicroClime Environmental Lid to Reduce Edge Effects in a Cell-based Proliferation Assay
- Folletos, volantes y fichas técnicas
-
Estudio de Casos
- Adenoviral Vectors Preparation
- Algae Biofuel Production
- Autophagy
- B Cell Research
- Basic Research on Reproductive Biology
- Cardiovascular Disease Research
- Cell Marker Analysis
- Choosing a Tabletop Centrifuge
- Collagen Disease Treatment
- Controlling Immune Response
- Creating Therapeutic Agents
- DNA Extraction from FFPE Tissue
- English Safety Seminar
- Equipment Management
- Exosome Purification Separation
- Future of Fishing Immune Research
- Hematopoietic Tumor Cells
- iPS Cell Research
- Fundamentos de la purificación de proteínas de membrana para cristalografía de rayos X
- Organelles Simple Fractionation
- Particle Interaction
- Quality evaluation of gene therapy vector
- Retinal Cell Regeneration
- Sedimentary Geology
- Severe Liver Disease Treatment
- Treating Cirrhosis
- University Equipment Management
- Fundamentals of Ultracentrifugal Virus Purification
- Catálogos
- Libros electrónicos
- Folletos
- Entrevistas
-
Posters
- AMP 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Applications of Ultracentrifugation in Purification and Characterization of Biomolecules
- Automating Genomic DNA Extraction from Whole Blood and Serum with GenFind V3 on the Biomek i7 Hybrid Genomic Workstation
- ABRF 2019: Automated Genomic DNA Extraction from Large Volume Whole Blood
- Automated library preparation for the MCI Advantage Cancer Panel at Miami Cancer Institute utilizing the Beckman Coulter Biomek i5 Span-8 NGS Workstation
- Automating Cell Line Development for Biologics
- Cellular Challenges: Taking an Aim at Cancer
- Cell-Line Engineering
- Characterizing the Light-Scatter Sensitivity of the CytoFLEX Flow Cytometer
- ASHG 2019: Comparison between Mutation Profiles of Paired Whole Blood and cfDNA Samples
- ASHG 2019: Correlation Between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- AACR 2019: Isolation and Separation of DNA and RNA from a Single Tissue or Cell Culture Sample
- Mastering Cell Counting
- AACR 2019: Correlation between Mutations Found in FFPE Tumor Tissue and Paired cfDNA Samples
- Preparing a CytoFLEX for Nanoscale Flow Cytometry
- A Prototype CytoFLEX for High-Sensitivity, Multiparametric Nanoparticle Analysis
- AGBT 2019: A Scalable and Automatable Method for the Extraction of cfDNA
- ABRF 2019: Simultaneous DNA and RNA Extraction from Formalin-Fixed Paraffin Embedded (FFPE) Tissue
- A Complete Automation and Reagent Workflow for Analysis of cfDNA: from Plasma to Variants
- Instrucciones del producto
- Protocolos
-
Informes técnicos
- La centrifugación es una solución de flujo de trabajo completa para la purificación de proteínas y la cuantificación de la agregación proteica
- AUC Insights - Analysis of Protein-Protein-Interactions by Analytical Ultracentrifugation
- GMP Cleanrooms Classification and Routine Environmental Monitoring
- AUC Insights - Assessing the quality of adeno-associated virus gene therapy vectors by sedimentation velocity analysis
- AUC Insights - Sample concentration in the Analytical Ultracentrifuge AUC and the relevance of AUC data for the mass of complexes, aggregation content and association constants
- Characterization of RNAdvance Viral XP RNA Extraction Kit using AccuPlex™ SARS–CoV–2 Reference Material Kit
- CytoFLEX Platform Flow Cytometers with IR Laser Configurations: Considerations for Red Emitting Dyes
- Hydraulic Particle Counter Sample Preparation
- Inactivation of COVID–19 Disease Virus SARS–CoV–2 with Beckman Coulter Viral RNA Extraction Lysis Buffers
- Liquid Biopsy Cancer Biomarkers – Current Status, Future Directions
- MET ONE 3400+ IT Implementation Guide
- SuperNova v428: New Bright Polymer Dye for Flow Cytometry
- Japan Document
-
Notas de aplicación